Thermal Holding Environmental Protection RC-60E Smoke and Dust Automatic Tester
Category:
Instrumentation/Gas analyzer/flue gas analyzer
Model:
RC-60E
Brand:
Thermal Holding Environmental Protection
model:
RC-60E
resolution:
1PPM
measurement object:
flue gas
measurement range:
5000ppm
power supply voltage:
two hundred and twenty
type:
portable
display mode:
electron
structural style :
handheld
Processing & Customization:
yes
gas flow rate:
five hundred
Specifications:
handheld
manufacturer:
Thermal Holding Environmental Protection
Retail Price
29,800.00USD
重量
kg
- Product Description
-
model RC-60E
resolution 1PPM
measurement object flue gas
measurement range 5000ppm
power supply voltage two hundred and twenty
type portable
display mode electron
structural style handheld
Processing & Customization yes
gas flow rate five hundred
Specifications handheld
manufacturer Thermal Holding Environmental Protection
Description :
The execution standard for the RC-60E automatic smoke and dust testing instrument for thermal holding environmental protection is:
JJG968-2002 Smoke Analyzer
HJ/T48-1999 Technical Conditions for Smoke and Dust Samplers
JJG680-2007 Smoke and Dust Tester
Enterprise standard: Q/0213LBJ 002-2016

Product Features:
1. Can measure parameters such as smoke, moisture content, flow rate, dynamic pressure, static pressure, smoke temperature, oil fume, and smoke (such as O2, SO2, NO, NO2, CO, CO2, H2S)
2. Equipped with ZD-24 DC power box, it can be used for both AC and DC
3. Equipped with a flue gas preprocessor for dust filtration, heating, and dehydration treatment, the data is more accurate
4. One million sets of data storage, RS232 connected to printer, USB memory makes data easier to analyze
5. One click sampling function
6. The drying cylinder is equipped with drying and filtering functions, enabling long-term monitoring
7. Chinese intelligent point and shoot machine operation interface, digital LED screen rotates 360 ° left and right for more convenient operation
8. Internal structural design such as power-off protection and password protection
9. Real time reading of parameters such as dynamic pressure, static pressure, ambient temperature, smoke temperature, rotational speed, flow rate, flow rate, moisture content, etc
10. Sampling system, imported electrochemistry, and modular design ensure the accuracy of testing
11. Equipped with a multifunctional sampling tube bracket for high-altitude hole sampling

Thermal Holding Environmental Protection RC-60E Smoke and Dust Automatic Tester Product Parameters:
parameter range
resolution
error
sampling flow
5~ 80 L/min
0.1 L/min
≤±2.5%
Flow control stability
≪ ± 2% (voltage varies between 180V and 250V, resistance varies within 3-6 kPa)
Smoke dynamic pressure
0~2000 Pa
1 Pa
≤±1.5%
Smoke static pressure
-30.00 ~+30.00 kPa
0.01 kPa
≤±2 %
Front pressure of flowmeter
-40.00 ~0 .00kPa
0.01 kPa
≤±2 %
Temperature before flowmeter
-30.0 ~ 150.0℃
0.1℃
≤±1.5℃
Smoke temperature
0 ~ 500℃
1℃
≤±3℃
atmospheric pressure
70~130kPa
0.1 kPa
≤±2.5 %
humidity ratio
(0~60)%
0.1%
≤±1.5%
O2 (optional)
0 ~ 30%
0.1%
Indication error: ≤± 2%;
Repeatability: ≤ 2%;
Response time: ≤ 90s;
Stability: The displayed value changes by ≤ 5% within 1 hour.
SO2 (optional)
0~5700 mg/m3
1 mg/m3
NO2 (optional)
0~200 mg/m3
1 mg/m3
NO (optional)
0~1300 mg/m3
1 mg/m3
CO (optional)
0~5000 mg/m3
1 mg/m3
CO2 (optional)
0-30%
0.01%
H2S (optional)
0-300 mg/m3
1 mg/m3
Sampling pump load capacity
≥ 50 L/min (when resistance is -20kPa)
Data storage capability
One million sets
Sampling volume
999999 .9 L
0.1 L
≤±2.5%
Overall dimensions
335×145×280 mm
Instrument noise
<65dB(A)
Total weight
About 8.0 kg
power consumption
<100 W
Thermal Holding Environmental Protection RC-60E Smoke and Dust Automatic Tester
AfterSalesService :
Key words:- Smoke and dust automatic testing instrument