measurement range:

0-100%

measurement object:

flammable gas

measurement accuracy:

one

resolution:

1ppm

power supply:

two hundred and twenty

Shell material:

one

Protection level:

one

Explosion-proof rating:

one

Processing & Customization:

No

weight:

one

manufacturer:

Huari

Retail Price

5,000.00USD


重量

kg

  • Product Description
  • brand

    RAE/Huarui

    model

    FGM-3300

    type

    fixed type

    measurement range

    0-100%

    measurement object

    flammable gas

    measurement accuracy

    one

    resolution

    1ppm

    power supply

    two hundred and twenty

    Shell material

    one

    Protection level

    one

    Explosion-proof rating

    one

    Processing & Customization

    No

    weight

    one

    manufacturer

    Huari

    Description :

    American Huarui RAE FGM-3300 combustible gas detector

    American Huarui FGM-3300 combustible gas detector:

      RAEAlert EC is a toxic gas detector that adopts modular design, intelligent sensor detection technology, integral explosion-proof (d) structure, and fixed installation method. The standard configuration includes a dot matrix LCD display, a three wire 4-20mA analog and RS485 digital signal output, and optional modules such as programmable switch output. Customized products are provided according to user needs, and functions such as output signal fine-tuning are also supported to facilitate system networking and maintenance. Can detect various toxic and harmful gases such as CO, H2S, O2, SO2, Cl2, Nspan, HCN, NO, NO2, Pspan, ClO2, ETO, etc.

      Huarui FGM-3300 product features:

      Intelligent EC sensor, using intrinsic safety technology, can support multi gas and multi range detection, and can provide customized products according to user needs. It can achieve sensor interchangeability, offline calibration, and zero point self calibration without the need for tools

      Intelligent temperature and zero compensation algorithms enable instruments to have better performance

      Has good selectivity, avoiding interference from other gases on the detected gas

      Multiple signal outputs, which can be easily connected to industrial control systems such as PLC/DCS, and can also be used as a standalone control

      Large dot matrix LCD display, supporting both Chinese and English interfaces

      No need to open the lid, infrared operation, single person maintenance

      Local alarm indication, integrated sound and light (optional)

      The instrument has over range and reverse polarity protection, which can avoid the danger caused by improper human operation

      Rich electrical interfaces for users to choose from

      Certified by ATEX, UL, CSA, etc., with international quality

      Huarui FGM-3300 Technical Parameters:

      Detection Principle

      electrochemistry

      sensor

      Smart Sensor

      sampling method

      diffusion

      working power supply

      9~30VDC

      power

      30mA@24VDC

      output signal

      4~20mA, Adjustable tuning

      load impedance

      600Ω

      Other outputs

      RS485 Modbus, Relay module (optional) high alarm, low alarm and fault 30VDC/5A, normally open/normally closed optional

      detection accuracy

      ±2%F.S.

      Response time (T90)

      <15s

      electrical interface

      3/4” NPT M

      Installation method

      2 "riser/wall mounted installation

      display mode

      128 × 96 dot matrix LCD, supporting graphic and text display

      1 LED status indicator light, 4 LED function indicator lights

      Protection level

      IP65

      ambient temperature

      -40 ℃~+70 ℃ (limit value)

      -40 ℃~+55 ℃ (typical value)

      ambient humidity

      15-95% RH (no condensation)

      environmental stress

      86kPa ~106kPa

      Overall dimensions

      182mm×195mm×107mm(HWD)

      Shell material

      aluminum alloy

      weight

      1.5kg

      ATEX

      Ex dia IIC T4 (under evidence collection)

      UL/CSA

      Class I, Div. 1, Groups BCD T4 (under evidence collection)

      IECEx

      Ex dia IIC T4 (under evidence collection)

      Explosion-proof certification

      Ex d ia IIC T6

      Huarui FGM-3300 standard configuration

      Detector host

      Operating manual

      Standard gas hood

      Install attachments

      Optional accessories

      Relay module

      Sound and light

      ?

      Standard gas

      Pressure reducing valve

      Install the attachment

      Wire connector

      Tighten the nut tightly

      U-shaped installation clip

      Install the backboard

      Standard gas hood

      Rain cover

      ? Blocker