Haoen temperature rapid change test box rapid temperature change circulation tank rapid temperature change test box 5 degrees/min
Category:
Instrumentation/Test chamber and climate environment equipment/Other test chambers and climate equipment
Model:
HE-GDK-225D9
Brand:
Haoen Instrument
model:
HE-GDK-225D9
outer box material:
Paint or stainless steel
inner material:
304 stainless steel
studio dimensions:
600*750*500
temperature range:
-70~+150
measurement range:
225 liters
Processing & Customization:
yes
Retail Price
1,000.00USD
重量
kg
- Product Description
-
model HE-GDK-225D9
outer box material Paint or stainless steel
inner material 304 stainless steel
studio dimensions 600*750*500
temperature range -70~+150
measurement range 225 liters
Processing & Customization yes
Description :
1Temperature rapid change test boxPerformance:
Refers to the air-cooled type at room temperature of 20 ℃ and when unloaded.
Model: HE-GDK-80/100/150/225/408/800/1000 (A, B, C, D)
Capacity: 80 liters, 100 liters, 150 liters, 225 liters, 408 liters, 800 liters, 1000 liters
Inner box size (width * height * depth): 400 X500X400、500X500X400、500X600X500、600X750X500、600X850X800、1000X1000X800、1000X1000X1000(MM)
Dimensions (width * height * depth): 980 X1590X940、1080X1590X940、1080X1690X1040、1080X1840X1130、1180X1950X1350、1600X2070X1400、1600X2070X1570(MM)
Temperature range: A represents: 0 ℃ to+150 ℃
B represents: -20 ℃ to+150 ℃
C represents: -40 ℃ to+150 ℃
D represents: -70 ℃ to+150 ℃
Temperature accuracy: ± 0.01 ℃
Temperature fluctuation: ± 0.5 ℃
Temperature uniformity: ± 2.0 ℃
Heating rate: 3 ℃/min, 5 ℃/min, 10 ℃/min (non-linear, unloaded); Non standard specifications can be customized according to requirements
Cooling rate: 3 ℃/min, 5 ℃/min, 10 ℃/min (non-linear, unloaded); Non standard specifications can be customized according to requirements
Power supply: three-phase five wire 380V 50HZ
Inner box material: SUS 304 mirror stainless steel
Outer box material: SUS 201 yarn surface stainless steel or cold steel plate spray treatment

IIRapid temperature change circulation tankControl system:
1. Adopting imported high-precision temperature and humidity control instruments, with touch screen control.
2. Temperature control adopts PID+SSR system for coordinated control of the same channel.
3. It has the function of automatic calculation, which can immediately correct the temperature change conditions, making temperature and humidity control more accurate and stable.
4. The real-time operation curve diagram of the controller operation interface can be displayed on the screen.
5. Adopt a temperature resistant and low-noise air conditioning motor and a multi blade centrifugal fan.

IIIRapid temperature change test chamberrefrigeration system
1. Compressor: Fully enclosed French Taikang compressor
2. Cooling method: single (double) mechanism cooling
3. Condensation method: forced air cooling
4. The solenoid valve, oil separator, dry filter, repair valve, and refrigerant flow window are all imported original parts

AfterSalesService :
Key words:- Temperature rapid change test box
More Products