DN300 large pipeline electromagnetic flowmeter
DN300 large pipeline electromagnetic flowmeter

DN300 large pipeline electromagnetic flowmeter

Category:

Instrumentation/Flow meter/electromagnetic flowmeter

Model:

LDG-300

Brand:

Jiangsu Fenghui Instrument Co., Ltd

Retail Price

2,800.00USD


重量

kg

  • Product Description
  • Description :

      DN300 large pipeline electromagnetic flowmeter

      The main components of this product are: measuring tube, electrode, excitation coil, iron core, and magnetic yoke housing. It is mainly used to measure the volumetric flow rate of conductive liquids and slurries in closed pipelines. Including highly corrosive liquids such as acids, alkalis, salts, etc.

      1、 Characteristics:

      The structure is simple, reliable, without moving parts, and has a long working life.

      No flow blocking components, no pressure loss or fluid blockage.

      No mechanical inertia, fast response, good stability, can be applied to automatic detection, adjustment, and program control systems.

      The measurement accuracy is not affected by the type of measured medium and its physical parameters such as temperature, viscosity, density, pressure, etc.

      Different combinations of PTFE or rubber lining materials and electrode materials such as HC, HB, 316L, Ti can adapt to the needs of different media.

      There are various flow meter models available, including pipeline type and plug-in type.

      Using EEPROM memory, the storage of measurement and operation data ensures safety and reliability.

      It has two forms: integrated and separated.

      High definition LCD backlit display.

      DN300 large pipeline electromagnetic flowmeter

      2、 Technical parameters:

      Nominal diameter (mm) (special specifications can be customized)

      Pipeline type PTFE lining: DN10~DN600

      Pipeline rubber lining: DN40~DN2200

      Flow direction: positive, negative, net flow

      Range ratio: 150:1

      Repetitive error: ± 0.1% of the measured value

      Accuracy level: Pipeline type: 0.5 level, 1.0 level

      Temperature of the tested medium: ordinary rubber lining: -20 to+60 ℃; High temperature rubber lining: -20 to+90 ℃

      PTFE lining: -30 to+100 ℃; High temperature PTFE lining: -20 to+180 ℃

      Rated working pressure: (high pressure can be customized) DN10- DN150: ≤ 1.6MPa; DN200-DN350:≤1.0MPa

      DN400-DN1000:≤0.6MPa; DN450-DN2200:≤0.25MPa

      Flow velocity range: 0.1-15m/s

      Conductivity range: The conductivity of the measured fluid is ≥ 5 μ s/cm

      Signal output: 4-20mA (load resistance 0-750 Ω), pulse/frequency, control level

      Communication output: RS485, MODBUS protocol, HART protocol

      Power supply: AC220V or DC24V

      Requirement for straight pipe length: upstream ≥ 5DN, downstream ≥ 2DN

      DN300 large pipeline electromagnetic flowmeter

      3、 Selection:

      The selection of instruments is a very important task in instrument applications. According to relevant data, 2/3 of instrument failures in practical applications are caused by incorrect selection and installation of instruments, and special attention should be paid.

      collect data

      ① Name of the tested fluid;

      ② Flow range;

      ③ Work pressure;

      ④ Working temperature.

      The tested fluid must have a certain conductivity, with a conductivity of ≥ 5uS/cm.

      Traffic and minimum traffic.

      The actual working pressure must be less than the rated working pressure of the flowmeter.

      The working temperature must meet the temperature requirements specified by the flowmeter.

      Determine if there is negative pressure present.

    AfterSalesService :

    Key words:
    • DN300 electromagnetic flowmeter