LDG-50 carbon steel electromagnetic flowmeter
This product is an electromagnetic induction flowmeter with a nominal diameter range of 10 to 3000. It is a series of electromagnetic flowmeter products that integrate intelligence, small and lightweight integration, multifunctionality, high precision, and high reliability.
1、 Technical performance indicators:
Flow velocity measurement range: (0.05-10) m/s
Power consumption:<15VA
Protection level: IP68
Connection method: flange connection
Electrode materials: molybdenum containing stainless steel OCr18Ni12Mo2Ti, Hastelloy B, Hastelloy C, titanium Ti, tantalum, etc
Measuring pipeline diameter: DN (32-6000) mm
Instrument accuracy: Level 1.0
Work pressure: ≤ 1.6MPa
Medium temperature: 0-80 ℃

LDG-50 carbon steel electromagnetic flowmeter
2、 Functional features:
1. Adopting programmable low-frequency rectangular excitation waves with good stability;
2. Adopting a 16 bit embedded microprocessor, it has fast computing speed and high accuracy;
3. Adopting full digital processing, it has strong anti-interference ability, stable performance, and a measurement range of up to 150:1
4. Adopting high-definition LCD backlight LCD display;
5. It has bidirectional flow measurement and bidirectional total accumulation functions, and bidirectional output functions for current and frequency;
6. Optional M-bus and HART protocols;
7. Equipped with self checking and self diagnostic functions;
8. Widely used in industrial technology and management departments such as chemical industry, environmental protection, metallurgy, medicine, papermaking, water supply and drainage.
3、 Installation:
1. Determine the installation location in order to make a reasonable choice between integrated or split electromagnetic flow meters;
2. Clearly measure the diameter of the measured pipeline;
3. Clarify whether the medium is conductive; (Conductivity>5 μ S/cm)
4. Clearly define their work pressure, working temperature, and working flow rate;
5. Clearly select a lining and electrode material with reasonable chemical properties as the medium.