Dual power supply DN80 gas turbine flowmeter DN80 intelligent gas turbine flowmeter
Category:
Instrumentation/Flow meter/flowmeter
Model:
LWQ-80
Brand:
Fenghui Instrument
Retail Price
5,200.00USD
重量
kg
- Product Description
-
Description :
This product has the advantages of high precision, good repeatability, simple structure, high pressure resistance, wide measurement range, small size, light weight, low pressure loss, long service life, simple operation, and convenient maintenance. It is a temperature and pressure compensation instrument.
1、 Technical parameters:
1. Instrument caliber and connection method: 25, 40, 50, 80, 100, 150, 200, 250 adopt flange connection; 25 and 40 can be connected by threads;
2. Accuracy level: ± 1.5% R, ± 1% R;
3. Range ratio: 1:10; 1:20; 1:30;
4. Display mode: Dual row LCD displays both instantaneous flow and cumulative flow simultaneously;
5. Instrument material: Body: 304 stainless steel; Impeller: anti-corrosion ABS or high-quality aluminum alloy; Display: cast aluminum;
6. Temperature of the measured medium (℃): -30 ℃ to+80 ℃;
7. Environmental conditions: Medium temperature: -30 ℃~+80 ℃, relative humidity: 5%~90%, atmospheric pressure: 86~106Kpa
8. Output signal: Standard two-wire 4-20mA (LWGQ-C type);
9. Transmission distance: ≤ 1000m;
10. Explosion proof grade: ExdIIBT6.

2、 Characteristics:
1. High quality alloy turbine with higher flow stability and corrosion resistance;
2. Imported high-quality specialized bearings with long service life;
3. The isolation between the measuring room and the ventilation room ensures the safety of the instruments;
4. It can detect the temperature, pressure, and flow rate of the tested gas, automatically track and compensate for the flow rate, and display the standard state; The cumulative gas volume of Pb=101.325KPa, Tb=293.15K; Real time query of temperature and pressure values;
5. Wide flow range (Qmax/Qmin ≥ 20:1), good repeatability, high accuracy (up to 1.0 level), low pressure loss, low starting flow rate, up to 0.6m3/h;
6. Multi point nonlinear correction of intelligent instrument coefficients;
7. Built in pressure and temperature sensors, high safety performance, compact structure, and beautiful appearance;
8. The instrument has explosion-proof and protective functions, with explosion-proof markings of Exd Ⅱ BT6 and Exia Ⅱ CT6, and a protection level of IP65.
3、 Installation:
Sensors can be installed in vertical, horizontal, or any inclined position according to the flow direction indicator;
When the pipeline is long or close to the vibration source, supports should be installed upstream and downstream of the flowmeter to eliminate the impact of pipeline vibration;
The installation location of the sensor should have sufficient space for the inspection and maintenance of the flowmeter, and should meet the environmental requirements of the flowmeter;
Avoid interference from strong external magnetic fields;
When installed and used outdoors, there should be a cover to avoid direct sunlight and rainwater erosion, which may affect the service life of the instrument;
When conducting pipeline pressure testing, attention should be paid to the pressure measurement range of the pressure sensor equipped on the intelligent flowmeter to avoid damaging the pressure sensor due to overpressure.
Attention should be paid to the influence of installation stress. The upstream and downstream pipelines of the flowmeter should be coaxial, otherwise shear stress will be generated.
The installation location of the flowmeter should consider the thickness of the sealing gasket or install an elastic expansion joint on the downstream side.
Before installing the flow meter, the welding slag and other debris in the pipeline should be removed first.

AfterSalesService :
Key words:- Dual power supply DN80 gas turbine flowmeter