- Product Description
-
Description :
This product can measure the flow rate of dirty media, corrosive media, and suspended liquid-solid two-phase flow. This is because there are no flow obstructing components inside the instrument measuring tube, and only the inner lining and electrodes of the measuring tube are in contact with the measured fluid. The material can be selected according to the properties of the measured fluid.
1、 Characteristics:
There are no obstruction components inside the flowmeter, and there is almost no pressure loss or fluid blockage.
No mechanical inertia, fast response, wide fluid measurement range (flow velocity 0.3-12m/s), good stability, can be used for automatic detection, adjustment, and programmable systems.
Measuring liquids with conductivity greater than 5u S/cm, the measurement is not affected by changes in fluid density, viscosity, temperature, pressure, and conductivity. The sensor's induced voltage signal is linearly related to the average flow rate, and the measurement accuracy is high.
Accuracy levels: 0.2, 0.5, 1.0, 1.5. Can meet the needs of different users.
The sensor part only has the inner lining and electrodes in contact with the measured liquid. As long as suitable electrode and inner lining materials are selected, it can be corrosion-resistant and wear-resistant.
Flow scoring for conventional type (pressure ≤ 4.0MPa) and high-pressure type (pressure ≥ 4.0MPa).
The flow meter system is divided into simple installation type and online installation type.

2、 Technical parameters:
1. Measurable pipe diameter: Pipeline PTFE lining: DN10~DN600 Pipeline rubber lining: DN40~DN1200 Note: Special specifications can be customized
2. Accuracy level: Pipeline type: 0.5 level, 1.0 level
3. Temperature of the tested medium:
Ordinary rubber lining: -20 to+60 ℃; High temperature rubber lining: -20 to+90 ℃
PTFE lining: -30 to+100 ℃; High temperature PTFE lining: -30 to+180 ℃
4. Rated working pressure:
DN10—DN65:2.5MPa
DN80—DN150:1.6MPa
DN200—DN1200:1.0MPa
5. Output signal and load resistance: 4-20mADC, 0-500 Ω
6. Electrode materials: including molybdenum stainless steel, titanium (Ti), tantalum (Ta), Hastelloy (HB), etc
7. Protection level: Submersible type: IP68, others: IP65
8. Power supply: 220VAC
9. Straight pipe length: Pipeline type: upstream ≥ 5DN, downstream ≥ 2DN
3、 Installation:
You should stay away from equipment with strong magnetic fields.
Avoid places with strong vibrations and corrosiveness.
Pipelines that do not allow flow interruption should be equipped with shut-off valves and bypass pipelines when installing flow meters.
There should also be a certain straight pipe section upstream of the flowmeter sensor, usually between 5D and 10D.
It can be installed horizontally, vertically, and diagonally, but it is necessary to ensure that the measuring tube is coaxial with the process pipeline and that the measuring tube is always filled with liquid.
The measuring tube, housing, shielded wires of the leads, and the pipelines at both ends of the sensor must be reliably grounded.